alt="{AELAB}{Adjustable Pipettes}-{AE-MB Series}"
AELAB Manual Normal Adjustable Single-channel Pipette
+ Lightweight, ergonomic, low force design
+ Digital display clearly reads volume setting
+ The pipettes cover volume range of 0.1μl to 10ml
+ Easy to calibrate and maintain with tool supplied
+ Design helps avoid repetitive strain injuries
+ Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
+ The low part is available for autoclaving
+ Manufactured from innovative materials
AE-MA01 AE-MA02 AE-MA03 AE-MA04 AE-MA05 AE-MA06 AE-MA07 AE-MA08 AE-MA09 AE-MA10 AE-MA11
1
Quantity:

AELAB Manual Normal Adjustable Single-channel Pipette AE-MA01 AE-MA02 AE-MA03 AE-MA04 AE-MA05 AE-MA06 AE-MA07 AE-MA08 AE-MA09 AE-MA10 AE-MA11

Mechanical Pipettes (Adjustable & Fixed)
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves

gravimeter testing of each pipette with distilled water at 22℃  .

Manual Pipette Type
+ Normal Adjustable/Fixed Manual 1,8,12 Channels Pipette (Non-Sterile)
+ Half Autoclave Adjustable/Fixed Manual 1,8,12 Channels Pipette
+ Fully Autoclave Adjustable/Fixed Manual 1,8,12 Channels Pipette

Features for Single Channel
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
- Manufactured from innovative materials


Features for Muti-channels

- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips.



Technical Specifications
Single-channel Adjustable Volume Pipettes
PictureModelVolume range IncrementTest Volume Maximum permissibleMaximum Permissible
systematic error random error
(Inaccuracy)(Imprecision)
(μl)%μl%μl
AE-MA010.1-2.5μL0.05μL2.5μL2.50%0.06252.00%0.05
1.25μL3.00%0.03753.00%0.0375
0.25μL12.00%0.036.00%0.015
AE-MA020.5-10μL0.1μL10μL1.00%0.10.80%0.08
5μL1.50%0.0751.50%0.075
1μL2.50%0.0251.50%0.015
AE-MA032-20μL0.5μL20μL0.90%0.180.40%0.08
10μL1.20%0.121.00%0.1
2μL3.00%0.062.00%0.04
AE-MA045-50μL0.5μL50μL0.60%0.30.30%0.15
25μL0.90%0.2250.60%0.15
5μL2.00%0.12.00%0.1
AE-MA0510-100μL1μL100μL0.80%0.80.15%0.15
50μL1.00%0.50.40%0.2
10μL3.00%0.31.50%0.15
AE-MA0620-200μL1μL200μL0.60%1.20.15%0.3
100μL0.80%0.80.30%0.3
20μL3.00%0.61.00%0.2
AE-MA0750-200μL1μL200μL0.60%1.20.15%0.3
100μL0.80%0.80.30%0.3
50μL1.00%0.50.40%0.2
AE-MA08100-1000μL5μL1000μL0.60%60.20%2
500μL0.70%3.50.25%1.25
100μL2.00%20.70%0.7
AE-MA09200-1000μL5μL1000μL0.60%60.20%2
500μL0.70%3.50.25%1.25
200μL0.90%1.80.30%0.6
AE-MA101000-5000μL50μL5000μL0.50%250.15%7.5
2500μL0.60%150.30%7.5
1000μL0.70%70.30%3
AE-MA112-10mL0.1mL10mL0.60%600.20%20
5mL1.20%600.30%15
2mL3.00%600.60%12