TOC-1500/TOC-1700Total Organic Carbon Analyzer
Total Organic Carbon Analysis has been an analytical technique used to measure water quality in the water supply industry since in the early 1970s.
Different from TOC-2000 and TOC-3000, TOC-1500/1700 is equipped with a conductivity detector to quantify TOC.
TOC-1500 only works under off-line mode while TOC-1700 integrates on-line sampling mode which enables easy and quick sample introduction under onlinemode.
They are ideal choice to measure and monitor microelectronics water,purified water, water for injection, etc.
Features:
+ Automatic sample introduction with enter of trace sample, we can control the value of the trace sample. No contamination will during testing which will be harmful to human being and environment.
+ It adopts UV Catalytic oxidation during which there is no need to add acid, gas or catalytic with low lost of routine maintenance cost.
+ 7 inches touch screen with operation panel, which is easy to operate and maintain.
+ The system is set with password for protection which is in compliance with FDA-21 CFR Part11 Requirements. Also, it is in compliance with USP32-643 requirements.
+ Auto sampler is optional which can be used for testing various different samples without person.
+ Special designed for testing water less than 1.5 ppm. Online and office testing function can be easily switched.
+ 8GB mass storage memory which does not restrict date and time.
+ Less than 3 minutes for each analysis.
+ Clear historical records which is traceable according to analysis date.
+ Data for specific whole day can be derived and saved to USB directly.
+ Equipped with Bluetooth printer which is easy to operate and print quickly.
+ Modular design ensures quick installation and easy maintenance.
+ Optional online monitoring device meets different requirements.
Technical Specifications:
Model | AE-TOC-1500 | AE-TOC-1700 (online) |
Detection range | 0.001-1.500mg/l (ppm) |
Detection limit | 1ppb |
Max tolerance | ±5% |
Analysis time | 3 min |
Response time | within 10 min |
Sample temperature | 1-95℃ |
Environment temperature | 10-40℃ with temperature change ±5 ℃/d |
Sample flow speed | 3ml/min |
Relative humidity | ≤85% |
Repeatability tolerance | ≤3% |
Drifting | ±5% |
Power | 220/110V±10% |
Frequency | AC 220V/110V, 50/60Hz |
Power | 100W |
Dimension | 440*220*300mm |